Single direction, single axis units. Generally, start the selection with the maximum g amplitude range in mind. Next, verify other factors including temperature, range, mounting type, and cable configuration. Or browse the Shock, Miniature, General Purpose, Modal Array, or Industrial accelerometer sections. For multi-axis testing, consider a triaxial unit or use a triaxial mounting adaptor(see accessories). See calibration notes for information on calibration. Contact us for detailed selection assistance, or for unique requirements not shown.
Designed to withstand and measure extreme high-level, short-duration transient accelerations, units also feature internal filtering for resonance suppression. Applications include metal-to-metal tests, impacts and armor testing, explosive studies, pile driver monitoring, and simulated pyroshock events. High-g shock calibration is included with rental of 350 Series units.
10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
Rental Includes: Calibration - NIST traceable and ISO 17025/A2LA accredited calibration; includes 10g frequency response to upper 1dB point (TMS: MCS-A001 or PCB ACS-22) and high g shock (TMS: MCS-31 or PCB: ACS-14). See notes on calibration for additional information
Similar to: B&K 8339-002, Dytran 3200B, Kistler 8743A100
10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
Rental Includes: Calibration - NIST traceable and ISO 17025/A2LA accredited calibration; includes 10g frequency response to upper 1dB point (TMS: MCS-A001 or PCB ACS-22) and high g shock (TMS: MCS-31 or PCB: ACS-14). See notes on calibration for additional information
Similar to: B&K 8339-001, Dytran 3200B2, Endevco 2255B & 7255A, Kistler 8743A50 & 8742A50
10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
Rental Includes: Calibration - NIST traceable and ISO 17025/A2LA accredited calibration; includes 10g frequency response to upper 1dB point (TMS: MCS-A001 or PCB ACS-22) and high g shock (TMS: MCS-31 or PCB: ACS-14). See notes on calibration for additional information
Similar to: B&K 8339-001, Dytran 3200B2, Endevco 2255B & 7255A, Kistler 8743A50 & 8742A50
10-32 coaxial jack electrical connector used with 002C or 003C series cables
Rental Includes: Calibration - NIST traceable and ISO 17025/A2LA accredited calibration; includes 10g frequency response to upper 1dB point (TMS: MCS-A001 or PCB ACS-22) and high g shock (TMS: MCS-31 or PCB: ACS-14). See notes on calibration for additional information
Similar to: Dytran 3200B5, Endevco 2255B-1 & 7255A-1, Kistler 8743A5 & 8742A5
10-32 coaxial jack electrical connector used with 002C or 003C series cables
Rental Includes: Calibration - NIST traceable and ISO 17025/A2LA accredited calibration; includes 10g frequency response to upper 1dB point (TMS: MCS-A001 or PCB ACS-22) and high g shock (TMS: MCS-31 or PCB: ACS-14). See notes on calibration for additional information
Similar to: Dytran 3200B5, Endevco 2255B-1 & 7255A-1, Kistler 8743A5 & 8742A5
10-32 coaxial jack electrical connector used with 002C or 003C series cable
Rental Includes: 080A04-Adhesive mounting base, 0.37" hex x 0.2" thk, 10-32 tapped hole, aluminum w/insulating hardcoat finish 081B05-10-32 to 10-32 adaptor stud Calibration - NIST traceable and ISO 17025/A2LA accredited calibration; includes 10g frequency response to upper 1dB point (TMS: MCS-A001 or PCB ACS-22) and high g shock (TMS: MCS-31 or PCB: ACS-14). See notes on calibration for additional information
Lightweight units are ideal for high frequency testing with a small size and low mass. Applications include environmental testing, component qualification, structural testing and operational behavior of lightweight or thin objects, and fatigue testing.
! When removing teardrop style sensors, it is essential to use the removal tool supplied along with an appropriate debonding agent.
3-56 coaxial jack electrical connector. Included 10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
/NC (No Cable) option available-use your own or choose custom length and configuration here.
Technique video for installation and removal of miniature accelerometers is key to obtaining valid measurements and to avoid damage to the accelerometer.
Rental Includes: 030A10-10-ft detachable cable. 3-56 plug to 10-32 plug 039A27-Removal tool Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4516-001, Endevco 25A, 25B, 7625A & 2250A-10
Technique video for installation and removal of miniature accelerometers is key to obtaining valid measurements and to avoid damage to the accelerometer.
Rental Includes: 039A26-Removal tool Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
3-56 coaxial jack electrical connector. Included 10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
/NC (No Cable) option available-use your own or choose custom length and configuration here.
Technique video for installation and removal of miniature accelerometers is key to obtaining valid measurements and to avoid damage to the accelerometer.
Rental Includes: 030A10-10-ft detachable cable. 3-56 plug to 10-32 plug 039A26-Removal tool Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4516-001, Endevco 25A & 25B & 7625A
3-56 coaxial jack electrical connector. Included 10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
/NC (No Cable) option available-use your own or choose custom length and configuration here.
Technique video for installation and removal of miniature accelerometers is key to obtaining valid measurements and to avoid damage to the accelerometer.
Rental Includes: 030A10-10-ft detachable cable. 3-56 plug to 10-32 plug 039A27-Removal tool Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4516-001, Endevco 25A & 25B & 7625A
3-56 coaxial jack electrical connector. Included 10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
/NC (No Cable) option available-use your own or choose custom length and configuration here.
Technique video for installation and removal of miniature accelerometers is key to obtaining valid measurements and to avoid damage to the accelerometer.
Rental Includes: 030A10-10-ft detachable cable. 3-56 plug to 10-32 plug 039A26-Removal tool Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4517, Dytran 3224A1, Endevco 27A11 & 27AM1-10 & 2250A & 2222C (charge type), Kistler 8778A500
3-56 coaxial jack electrical connector. Included 10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
/NC (No Cable) option available-use your own or choose custom length and configuration here.
Technique video for installation and removal of miniature accelerometers is key to obtaining valid measurements and to avoid damage to the accelerometer.
Rental Includes: 030A10-10-ft detachable cable. 3-56 plug to 10-32 plug 039A26-Removal tool Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: Dytran 3225M24, Endevco 27A12 & 27AM1-100 & 2222D (charge type)
These units cover a wide range of tests ranging from vibration control to machine testing, and product qualification. Sensors detailed measure either ±50g or ±500g ranges, but piezoelectric accelerometers often measure an extremely high dynamic range and are perfectly suited for measurements of quite low vibration levels.
5-44 coaxial jack electrical connector used with 003P or 018C cable series
Rental Includes: Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from 5 Hz to 10 kHz. Data to upper 3dB point included using resonance sweep or gravimetric method; high frequency calibration is traceable but not A2LA accredited.
5-44 coaxial jack electrical connector used with 003P or 018C cable series
Rental Includes: 080A15-Adhesive mounting base, 0.312" hex x 0.125" thk, 5-40 tapped hole, aluminum w/insulating hardcoat finish Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4519, Dytran 3035B1, Endevco 256-10, Kistler 8720A500
10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
Rental Includes: 080A15-Adhesive mounting base, 0.312" hex x 0.125" thk, 5-40 tapped hole, aluminum w/insulating hardcoat finish Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A15-Adhesive mounting base, 0.312" hex x 0.125" thk, 5-40 tapped hole, aluminum w/insulating hardcoat finish Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4518, Dytran 3145A1, Kistler 8730A500
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A-Adhesive mounting base, 0.5" hex x 0.187" thk, 10-32 tapped hole, aluminum w/insulating hardcoat finish 081B05-10-32 to 10-32 adaptor stud Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A-Adhesive mounting base, 0.5" hex x 0.187" thk, 10-32 tapped hole, aluminum w/insulating hardcoat finish 081B05-10-32 to 10-32 adaptor stud Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A15-Adhesive mounting base 081B05-Mounting Stud (10-32 to 10-32) Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A-Adhesive mounting base 081B05-Mounting Stud (10-32 to 10-32) Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
5-44 coaxial jack electrical connector used with 003P or 018C cable series
Rental Includes: 080A15-Adhesive mounting base, 0.5" hex x 0.187" thk, 10-32 tapped hole, aluminum w/insulating hardcoat finish Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4519-001, Dytran 3035B, Endevco 256-100, Kistler 8776A50
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A15-Adhesive mounting base, 0.5" hex x 0.187" thk, 10-32 tapped hole, aluminum w/insulating hardcoat finish Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4518-001, Dytran 3145A, Kistler 8774A50
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A15-Adhesive mounting base, 0.5" hex x 0.187" thk, 10-32 tapped hole, aluminum w/insulating hardcoat finish Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A15-Adhesive mounting base, 0.5" hex x 0.187" thk, 10-32 tapped hole, aluminum w/insulating hardcoat finish Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4518-001, Dytran 3145A, Kistler 8774A50
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 081B05-10-32 to 10-32 adaptor stud Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Excellent phase characteristics and lightweight construction, ideal for structural vibration, multi-channel modal analysis, analytical model correlation, and force-response testing. Many units have TEDS functionality as an option, see information below. A wide range of structural test hardware is also available, including shakers and impact hammers, as well as large-channel cable management solutions.
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4507-002, Dytran 3097A2, Kistler 8640A50
3 pin socket connector used with 080B cable series
Rental Includes: Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A25-Adhesive mounting base, 0.438" hex, 5-40 tapped hole, aluminum hardcoat Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
10-32 coaxial jack electrical connector used with 002C or 003C cable series
Rental Includes: 080A25-Adhesive mounting base, 0.438" hex, 5-40 tapped hole, aluminum hardcoat Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Adjusted circuitry allows measurement of these units in temperatures to +325°F (163°C) and are used with traditional ICP power systems. Slight low frequency and broadband resolution specifications may be affected.
High temperature version of 352A21, temperature range -65 to +325 °F
0.45"x0.25"x0.14" Titanium sensor
±500g range
0.6 gram mass
Adhesive mounting
3-56 coaxial jack electrical connector. Included 10-ft integral cable terminating in 10-32 plug. Use 070A02 scope adaptor to terminate in BNC, or use 070A05 cable connector to 002C or 003C series cable.
/NC (No Cable) option available-use your own or choose custom length and configuration here.
Technique video for installation and removal of miniature accelerometers is key to obtaining valid measurements and to avoid damage to the accelerometer.
Rental Includes: 030A10-10-ft detachable cable. 3-56 plug to 10-32 plug 039A27-Removal tool Calibration - NIST traceable and ISO 17025/A2LA accredited calibration from at least 10 Hz to upper 5% point (TMS: MCS-A001 or PCB: ACS-1). See notes on calibration for additional information
Similar to: B&K 4516-001, Endevco 25A & 25B & 7625A
Calibration data is provided electronically (on CD or USB memory stick) or as printed copies, depending on quantity rented and end-user preference.
The Modal Shop has internal calibration intervals set for most test products at 1 year. ANSI/ISO/IEC 17025:2000 makes clear that it is the responsibility of the end-user organization (in this case, the renter) to determine the appropriate calibration interval under the requirements of its own quality system. Per this, calibration certificates of rented equipment have "calibrated on" dates, but blank "calibration due" dates. Rentals ship with a customer specific expected return date in mind - if equipment is expected out for 5 months, the last calibration date will be within the past 7 months. Calibration to additional standards (e.g. Z540 or specific FAA requirements) are possible, but there may be an additional charge.
The Modal Shop's Scope of Accreditation can be found here. An example of a calibrations certificate can be found here. Please contact us with questions or if a different calibration will be needed for your rental sensors.
Notes on TEDS (Transducer Electronic Data Sheet) Back to Top
The TEDS option is available on several rental transducers. TEDS stands for Transducer Electronic Data Sheet and provides memory within a sensor with manufacturer name, sensor type, model number, serial number and calibration data. It is an approved IEEE standard (1451.4), and many data acquisition manufacturers include the ability to read TEDS sensors. Not sure if your DAQ has TEDS capability? No problem; TEDS sensors can be used alongside non-TEDS sensors with no issues; the TEDS memory is bypassed. Want more information? See our TEDS FAQ .